A.S.I.LAB’s Class II, Type A2

A.S.I.LAB’s Class II, Type A2 Biosafety Cabinets provide personnel, product and environmental protection from hazardous particulates such as agents that require Biosafety Level 1, 2 or 3 containments. Other appropriate applications include work with antineoplastic drugs, genetic material, asbestos and additional substances that generate hazardous airborne particulates.
These biological safety cabinets are designed to discharge HEPA-filtered exhaust air directly into the laboratory, or into an exhaust system through the optional Canopy Connection. When canopyconnected, the Biosafety cabinets may be used for applications involving minute quantities of volatile toxic chemicals and tracer amounts of radionuclide as an adjunct to microbiological research.
-
Understanding BSC Class2 A2 works
During operation, room air is drawn into the inlet grill. The air in the plenum beneath the work surface is a mixture of unfiltered room air and air that has just passed through the work area.
​
This contaminated air is drawn by the blower through the back plenum of the cabinet, where approximately 70% of the air is recalculated through the supply HEPA filter and back over the work area. The balance of the contaminated air is discharged to the environment after passing through the exhaust HEPA filter.
-
Features to be noticed
Available in various internal width models as follows
-
ASBSC-2 A2 – 2ft width
-
ASBSC-3 A2 - 3ft width
-
ASBSC-4 A2 – 4ft width
-
Other customizable models with longer widths.
-
10° Angled sash with Motorized opening and closing.
-
Microprocessor-controlled logic and LCD display.
-
Exterior body made of high strength mild steel duly epoxy coated.
-
Inner chambers made of 304grade stainless steel.
-
Easily transferable work surface Tray providing thorough decontamination of tray and internal chamber of unit.
-
Nominal inflow velocity of 100 fpm ±10fpm(0.5 m/sec)
-
Nominal down flow velocity of 55 fpm±10fpm (0.3 m/sec).
-
Approximately 70% recirculation and 30% exhaust.
-
HEPA filters- Two HEPA filters (Exhaust and supply), should be of 99.9% efficiency at 0.1to 0.3μ.
-
Motorized sash opening and closing with microprocessor-controlled logic and LCD display.
-
UV Lamp. For secondary decontamination while cabinets not in use.
-
Safety switch permits operation only when sash is completely closed
-
Fluorescent light for better visibility in the working area.
-
Blower speed automatically reduces when the sash is completely closed.
-
Sash opening available in two options 8 inches or 10 inches as per user requirement**
-
Audio visual alarm for Sash height; UV light hours, filter replacement; (optional Airflow slow)
-
5mm tempered glass sash / polycarbonate sash**.
-
2 electrical sockets of 230v/50hz 15amp and 6 amp plugs *
-
Optional Features
-
Service fixture or Ball valve for supplying water, Vacuum, Air or noble gases to cabinet interior.
-
Base stand.
-
IV stand or hanger rod.
-
Drainage Tray
-
Prior to ordering
-
Decide the model no.
-
Decide the required optional feature.
-
This part can be increased as per customer requirement
-
To be finalized prior to ordering